Probable Cause Actions
Error in performing Read the instructions
the test. thoroughly and repeat
the test again.
Control solution bottle Shake the control
not shaken well. solution bottle vigorously
and repeat the test again.
Expired or Check the expiration
contaminated date and the discard date
control solution. of the control solution.
Control solution that Control solution
is too warm or too cold.
should come
to room temperature
(less than 86°F/30°C)
before testing.
Test strip deterioration. Please repeat the test
with a new test strip.
v The control solution test is out of
range.
62
Probable Cause Actions
Meter malfunction. Please contact
Customer Care at
1.800.243.2636 for
assistance.
v The control solution test is out of
range.
63