STEP 4: Check if the Test Result is in Range.
Inlessthan(7)seconds,thecontrol
solution test result appears on the display.
Compare the test result with the range
printed on the test strip vial. The test
result should fall within the printed range.
See Figure 23.
Youwillhearavoicemessage,
“Blood glucose is
(number) (unit)”.
Out of Range Results
If test results fall outside the range printed
on the test strip vial, check “Trouble-Shooting
Guide” located in the “System Troubleshooting”
Sectionandrepeatthetest.
If you continue to get out-of-range results, it means that the
system or the control solution may not be working properly.
DO NOT use the system to test your blood glucose level. If
you are unable to resolve the problem, contact Technical
Supportat1-800-243-2636.
Figure 23
STEP 3: Apply Control Solution
Shakethecontrolsolutionbottle
well.Removecap.Squeezethe
bottle, discard the first drop, and
wipe off the dispenser tip with a
clean tissue paper or cotton.
Squeezethebottleagaintogeta
second drop onto a clean
non-absorbent surface or on your
fingertip first. Then apply the drop to
the opening of the absorbent channel
oftheteststrip(whereitmeetsthe
narrowchannel)untiltheconrmation
window is filled. The meter begins
to count down. See Figures 20, 21 and 22.
Caution: To avoid contaminating the control solution with
the content of the test strip, DO NOT DIRECTLY APPLY
CONTROL SOLUTION ONTO THE TEST STRIP.